Fiber-Coupled SureLock 785nm/830nm/1064nm Butterfly Raman Laser SureLock 785nm/830nm/1064nm Butterfly Raman Laser
Fiber-Coupled SureLock 785nm/830nm/1064nm Butterfly Raman Laser

SureLock™ 785 nm/830 nm/976 nm/ 1064 nm Butterfly Raman Lasers

Coherent’s fiber-coupled and free-space butterfly lasers are designed specifically for Raman spectroscopy applications, and offer high-power, wavelength-stabilized performance in both fiber-coupled (FC) and free space (FS) configurations. The narrowed linewidth, low power consumption, and broad stabilized temperature operating characteristics deliver affordable, portable instrument-quality performance in a broad choice of wavelengths from 785 nm to 1064 nm.

All SureLock™ Series lasers are stabilized using the Coherent PowerLocker® Volume Holographic Grating, ensuring precise, ultra-stable center wavelengths, low temperature dependence, and consistent optical performance over the locked region.

Fiber-coupled versions are available with FC/PC or FC/APC connectors.

Fiber-Coupled and Free-Space Butterfly Lasers

for Raman Spectroscopy

Specification Summary

Parameter Symbol Min Typ Max Unit
Output Power Po 600 mW
Center Wavelength(vacuum) LP 784.5, 829.5, 975.5, 1063.5 785, 830, 976, 1064 785.5, 830.5, 976.5, 1064.5 nm
Linewidth Δλ 0.07 0.15 nm
Central Stabilized Temperature1 Tc 20 40 °C
Stabilized Temperature Range1 Tr 14 °C

SureLock™ Specifications


Operating Specifications

Parameter Symbol Min Typ Max Unit
Threshold Current Ith
785/830nm
1064nm
325
250
mA
Operating Current Iop
785/830nm
1064nm
1100
1200
1500
1600
mA
Operating Voltage Vop
785/830nm
1064nm
1.9
2.1
2.2
2.5
V
TEC Current 2 A
TEC Voltage 4 V
Beam Divergence2 5 x 25 mrad
Beam Diameter2 0.2 (Vertical) x 1.0 (Horizontal) mm
Fiber Type3 105 µm core/ 900 µm tubing
Connector3 FC/PC (std)/ FC/APC (custom)
Numerical Aperture3 NA 0.22
Operating Temperature4 Top 0 25 50 °C
Storage Temperature4 Ts -20 80 °C

2 Free Space
3 Fiber Coupled
4 Non-condensing

Applications

Application

Raman Spectroscopy
Metrology
Bio-instrumentation
Sensing
Analytical Instrumentation

Features & Benefits


  • Compact 14-pin butterfly footprint
  • Narrow spectral linewidth 0.15 nm
  • Wavelength stability across operating range 0.01 nm/°C
  • Fiber-coupled output 105 µm MM fiber